Ammonia
Reaction Type (Mode) : | Fixed Time |
Reaction Direction : | Decreasing |
Wave Length : | 340 nm |
Flow Cell Temp. : | 37°C |
Zero Setting with : | Distilled Water |
Delay Time : | 40 Seconds (A1) |
Measuring Time : | 90 Seconds (A2) |
Reagent Volume: R1 | 300μl |
Reagent Volume: R2 | 200 μl |
Calibrator / Sample Volume : | 30 μl |
Calibrator Concentration : | 400 μg/dL |
Linearity : | 1700 μg/dL |
High Normal : | 150 μg/dL |