![Lipase 12 ml- 2](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-12-ml-2-1-scaled.jpg)
![Lipase 12 ml](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-12-ml-1-scaled.jpg)
![Lipase 50 ml - 5](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-50-ml-5-1-scaled.jpg)
![Lipase 50 ml](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-50-ml-1-scaled.jpg)
![Lipase 12 ml- 2](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-12-ml-2-1-150x150.jpg)
![Lipase 12 ml](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-12-ml-1-150x150.jpg)
![Lipase 50 ml - 5](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-50-ml-5-1-150x150.jpg)
![Lipase 50 ml](https://parikshabio.com/wp-content/uploads/2023/03/Lipase-50-ml-1-150x150.jpg)
Lipase
Reaction Type (Mode) | Fixed Time |
Reaction Direction | Increasing |
Wave Length | 578 nm |
Flow Cell Temp. | 37°C |
Zero Setting with | Distilled Water |
Delay Time | 5 Seconds |
Measuring Time | 120 Seconds |
Reagent Volume | 500 µL (R1) + 100 µl (R2) |
Calibrator / Sample Volume | 10 µl |
Calibrator Concentration : | 34 U/L |
Linearity | 300 |
Units | U/L |
High Normal | 64 U/L |